Useful stuff!

Wednesday, August 1, 2018

What ratio of isobaric labeling reagent to peptide is critical?


(Shoutout to A.J.Bureta for assembling part of this image and posting it under the Creative Commons Share-A-Like 3 where "free to modify" is rule number 2 {Hopefully my modifications aren't too obvious}).

Okay -- so I've noticed a trend recently -- I first noticed it right before ASMS in this great new paper. While at ASMS I hunted down a couple of studies that used isobaric labeling technologies and spoke to them about the ratio of peptides to labeling reagent they are using. The final straw was last week when I got to talk to a somewhat famous researcher in isobaric labeling techniques who referred me to some earlier works like this one

So...here is the question. When I pull up a protocol for isobaric labeling reagents, I'll find a recommended ratio of 8x labeling reagent to every 1x of peptides. In a protocol for another I find 10x.

Do we need that much reagent? In the first paper above, they clearly use a 2x labeling reagent to 1x peptides. In conversations with other researchers it sounds like 2x or 4x is becoming more of the secret (hidden deep in the methods section) norm. I can imagine that 8x probably does label closer to 100% of the peptides -- but if you could get 95% labeled with 2x or 97% with 4x  or 99% labeled with 8x --- umm..... that last 4%...ummmm --

[insert unrelated gif here]



....totally unrelated...


No comments:

Post a Comment